Test Procedures
- General checks will be made on the condition of any pre-filters within the workstation. The functions of the workstation will be checked for correct operation.
- Airflow readings are taken for all Powder Control Booths, measurements taken with a vane/hot wire anemometer, in accordance with BS EN ISO 14644-3 (2005).
- Filter integrity leak tests will be carried out on all HEPA filters (if fitted) in accordance with BS EN ISO 14644-3 (2005).
- Airborne particulate cleanliness counts will be carried out in accordance with BS EN 14644-1 (1999), and EU GMP Annex 1 (February 2008) (optional test).